000-00000-000-02-AXYZ Mapping TablesXY(Z)-150x150 / XY(Z)-200x150 / XYZ-150x150x100 / XYZ-200x150x100Installation and Operation ManualDocument Number
1: Setup2 000-10000-130-02-1209Unpacking the XYZ Table► Procedure1. Unpack your Mapping table carefully. Although the stage is rigidly mounted droppin
000-10000-130-02-1209 3Chapter 2XYZ Mapping Table SpecificationsThis section provides information on the operating environment, and electrical and mec
2: XYZ Mapping Table Specifications4 000-10000-130-02-1209Physical Specifications Specification Models XY-150x150 and XY-200x150 Models XYZ-150x150x10
000-10000-130-02-1209 5Chapter 3Operating InstructionsInterfacing with the XYZ Mapping TableThe XYZ Mapping Table can be operated by several different
3: Operating Instructions6 000-10000-130-02-1209To move the stage in the desired direction, press a mouse button over the appropriate arrow. The stage
3: Operating Instructions000-10000-130-02-1209 7Factory SettingsThe following settings can be reset in the controller by sending them to each controll
3: Operating Instructions8 000-10000-130-02-1209Axis Address node Example DescriptionX 1 1V200[CR] X-axis runs at constant speed of 100rpmY 2 2V300[CR
3: Operating Instructions000-10000-130-02-1209 9Commands for Motion ControlCommand Description ExampleM Initiate motionActivates positioning modeMLA L
3: Operating Instructions10 000-10000-130-02-1209Commands for Evaluating Limit Switches and Homing PointsCommand Description ExampleHO Define home pos
3: Operating Instructions000-10000-130-02-1209 11Commands for Evaluating Limit Switches and Homing Points (Cont’d)Command Description ExampleHB* Hard
Copyright © 2009 Ocean Optics, Inc.All rights reserved. No part of this publication may be reproduced, stored in a retrieval system, or transmitted, b
3: Operating Instructions12 000-10000-130-02-1209Commands for Parameters (Cont’d)Command Description ExampleAC* Load new accelerationParameter in Revo
3: Operating Instructions000-10000-130-02-1209 13Request CommandsCommand Description ExamplePOS Get actual position POS -> 5000GN Get actual speed
3: Operating Instructions14 000-10000-130-02-1209Request Commands (Cont’d)Command Description ExampleGAST Get actual status (4 Bits)Bit 0: 1 limit s
3: Operating Instructions000-10000-130-02-1209 15Request Commands (Cont’d)Command Description ExampleGES Get enhanced statusBit 0: 0Bit 1: 0 Bit 2:
3: Operating Instructions16 000-10000-130-02-1209
000-10000-130-02-1209 17Appendix AService and MaintenanceThough the Mapping Tables are mainly maintenance free, it may become necessary to lubricate t
A: Service and Maintenance18 000-10000-130-02-1209 z-axis:Drive the slide to center-position and bring in the tube of the injection to the drive scre
000-10000-130-02-1209 19IndexAASCII command set, 7Ccommandsbasic settings, 8homing points, 10limit switches, 10motion control, 9parameters, 11programm
Index20 000-10000-130-02-1209
000-10000-130-02-1209 AImportant Safety Notices1. All the safety and operating instructions should be read before the unit is operated.2. Dangerous vo
Important Safety NoticesB 000-10000-130-02-1209
000-10000-130-02-1209 iTable of ContentsAbout This Manual ...
000-10000-130-02-1209 iiiAbout This ManualDocument Purpose and Intended AudienceThis document provides you with an installation section to get your sy
About This Manualiv 000-10000-130-02-1209Authorization (RMA) number. Please contact an Ocean Optics Application Scientist for specific instructions wh
000-10000-130-02-1209 1Chapter 1SetupOverviewThe XY(Z) Mapping Table comes with an electric motor and encoder to drive each axis, along with a CNC con
Commentaires sur ces manuels